Installation, Inserting the module – Gilson 402 User Manual

Page 36

Advertising
background image

3-12

Installation

3

Inserting the Module

Do not try to force the module into position as this

may affect the alignment of the component parts.

You should proceed carefully as follows.

1) Tilt the top of the module inwards and carefully

insert the upper part of the module into the front

aperture. When the syringe drive motor is inside

the aperture, tilt the module into the upright

position.

2) Push the module into the aperture, making sure

that none of the cables become trapped between

the module and the edges of aperture. The module

rests on a rubber strip that dampens vibrations

and helps ensure the module is fitted correctly.

3) Secure the module in place at the top using the

two screws (provided).

4) Replace the cover by pulling the sides outwards

slightly and lowering the it over the 402. Insert

and tighten the four screws at the base of the 402

that secure the cover to the lower chassis. Do not

forget to reinstall the washers, which are designed

to ensure that the cover and chassis are con-

nected electrically.

5) Install the Tee-junction on the front of the

module using the two screws provided.

6) Install the syringe as described page 3-9.

7) Connect the hydraulic tubing according to pages

3-4 and 3-5.

8) Reconnect the power cord and switch the

instrument ON.

Upgrading to Dual-syringe Model