Technical data appendix, Junction tubing, Hydraulics – Gilson 402 User Manual

Page 112

Advertising
background image

Appendix C-4

Technical Data

Appendix

C

Hydraulics

Typical values (all syringes)

Syringe capacity Max. accuracy Max.error imprecision

%

+ A %

RSD %

100

0.50

0.10

50

0.80

0.20

10

1.00

0.50

The same specifications are obtained in the

dispensing mode with liquid from the reservoir.

See OQ Document for procedure and calculations.

Syringe

Volume Accuracy

RSD

capacity (µL) pipetted (µL)

%

µL

%

µL

100

1

5.0

0.05

3.0

0.03

250

2.5

3.0

0.075

2.0

0.05

500

5

3.0

0.15

2.0

0.1

1000

10

3.0

0.3

1.0

0.1

5000

50

3.0

1.5

0.8

0.4

10000

100

3.0

3

0.5

0.5

25000

250

3.0

7.5

0.5

1.25

For 1 % syringe capacity, typical

values are given.

Junction Tubing

(between valve and Tee)

Material: FEP, with two PVDF union connectors.

Tubing dimensions: 2.0 mm i.d., 70 mm length.

Tubing dead volume: 220 µL.

Advertising