5-9

Operation - Advanced Use

5

Counters

Under normal use you should expect to fit new (or

refurbished) syringes after 30000 cycles with water.

This feature is made easy by the 402, which automati-

cally counts the number of operating cycles performed

by the syringe(s) and valve. (A cycle is completed

when the syringe leaves its home position.) You

should only need to reset these counters after

changing one or more of these items. To do this,

press Count, the Syringe Counter Screen appears.
Press Chge, then key in the number from which

you wish the new count to start, followed by ENTER.

Press Next to go to the second syringe (if fitted) or

to the valve.
Press Chge, then key in the number from which the

new count is to start, followed by ENTER. Do this

for each counter that you want to change; followed

by ESC to return to the Configuration Screen.

Other Items

Press the Etc soft-key; the Beeper Screen appears.

This is to set an alarm when an error occurs, and to

warn that liquid is going to be dispensed to waste.
To switch between ON and OFF, press the Chge

soft-key, then press ENTER to confirm.
Press Next to go to the Autorefill Screen. When you set

Autorefill to ON, the left-hand syringe re-aspirates

diluent automatically for the next cycle, while you

are (for example) aspirating air via the probe. This

feature enables you to work faster with the 402.
To switch between ON and OFF, press the Chge

soft-key, then press ENTER to confirm. Press Prev

to go to the Beeper Screen. Press the ESC key to

return to the Configuration Screen.

Creating a Method (EDIT)

You use the keypad to enter the Method as a file

(one or more tasks used sequentially).

Key in the file number for your Method, then press

ENTER; press Edit. Now, one of two situations can

arise:
(i) if you have not entered a syringe correction factor

in the configuration process, the Task Selection

Screen appears, after which you press a soft-key

Software

Syringe Counter Screen

Beeper Screen

Autorefill Screen

Task Selection Screen