6-12

Maintenance

6

Changing the Piston Syringes and Piston Noses

Piston shaft

Piston
nose

Piston shaft holder

Piston nose holder

O-ring

5. Insert the new piston nose in the recess of the

piston nose holder. Make sure that the aperture is

facing outward and the O-ring is inside the seal.

6. Place the piston shaft in the piston shaft holder.
7. Screw the two parts of the tool together until tight.
8. Unscrew the tool, the piston nose should be

correctly seated on the shaft.

9. Dip the piston nose in distilled water and slide

the piston into the syringe.

10. Slide the piston up and down in the syringe five

or six times to ensure a smooth operation between

the piston and syringe.

11. Remount the syringe on the pump as described

page 6-4.