Operation - standard use, Essential steps – Gilson 402 User Manual

Page 48

Advertising
background image

4-12

Operation - Standard Use

4

Essential Steps

In Run Screen (1)

1

air gap

2

liquid

3

fraction1

4

fraction 2

5

fraction 3

6

fraction 4

7

to waste

Extra Volume

In Run Screen (2a)

In Run Screen (2b)

Ensure that the probe-end of the transfer-tubing is

in a suitable waste container, then press OK.

‘Homing ...’ appears on the screen until the

syringe(s) have returned to their home positions,

during which time any liquid contained in the

transfer-tubing is expelled from the probe-end.

After this the Action Screen reappears.
Press ESC to return to the start-up screen for the

current file.
Whenever you see ‘Homing’ on the screen, the

syringe piston(s) are returned to their uppermost

position(s). As a result, some (or all) of the liquid

contained in the transfer-tubing will be expelled

from the probe-end. You must ensure that the

probe end of transfer-tubing is already in a

suitable waste container.

Method Example

This section describes how to run Method 33

(created as above).
(i) Press Run, you see In Run Screen 1.

M33/T 1: DISPENS/1 means Method 33 Task 1,

Dispense Task, first cycle.
The cursor moves from side-to-side below the

parameter value field, indicating that you must

enter a value (or validate the current value, in

this example 6 mL/min.). Key in the required

flow rate (for example, 12 mL/min.) and press

ENTER. [Similar screens would appear for any

other parameters that you had entered as a

question mark (?).] The command light flashes

and you see alternately In Run Screens 2a and 2b

(about every half a second).
In the Dispense task, you aspirate an air gap, then

liquid. After aspirating, you dispense the liquid

in several fractions, finally the ‘Extra volume’ is

sent to waste.