Operation - advanced use – Gilson 402 User Manual

Page 63

Advertising
background image

5-13

Operation - Advanced Use

5

Press Abort, when you want to stop priming;

Initialization Screen 2 reappears. Ensure that the

probe-end of the transfer-tubing is in a suitable

waste container, then press OK. ‘Homing ...’

appears on the screen until the syringe(s) have

returned to their uppermost (home) positions,

during which time any liquid contained in the

transfer-tubing is expelled from the probe-end.

After this the Action Screen reappears.
Press ESC to return to the start-up screen for the

current file.
Whenever you see ‘Homing’ on the screen, the

syringe piston(s) are returned to their uppermost

position(s). As a result, some of the liquid con-

tained in the transfer-tubing will be expelled

from the probe-end. You must ensure that the

probe end of transfer-tubing is already in a

suitable waste container.

Measure

You can measure a volume (e.g. supernatant after

centrifugation).
• Press Measure, the Measure Screen appears.

For a Dual-syringe 402, press the Chge soft-key to

select LEFT or RIGHT. You are advised to select

the left-hand syringe if the volume to be measured

seems to be significantly larger than the capacity

of the right-hand syringe.
The maximum volume that you can measure is

limited to the volume of the aspirating syringe or

the transfer tubing size, whichever is the smaller.

• Press ENTER, the Flow Rate Screen appears.

Key in the flow rate at which you want the

volume measurement to take place, and then

press ENTER. The default flow rate is half the

maximum possible for the combination of syringe

and transfer tubing in use (refer to page 5-31).

Press Run, then follow the instructions given on the

screen; you press the command button on the probe

to aspirate, when prompted to do so by the com-

mand light on the probe:
- a steady light means you are to aspirate (or

dispense) liquid,

- a flashing light means you are to aspirate air via

the probe.

Software

Measure Screen

Flow Rate Screen