Maintenance, Changing the 500 µl piston noses – Gilson 402 User Manual

Page 93

Advertising
background image

6-9

Maintenance

6

Changing the Syringe Piston and Piston Noses

Changing the 500 µL Piston Noses

1. Remove the syringe from the pump as described

page 6-4.

2. Slide the piston out of the syringe barrel.
3. Remove the piston nose from the piston rod by

carefully pinching the piston nose with a pair of

cutters and gently pulling the piston nose off the

piston rod. Take care not to damage the end of

the piston rod when removing the piston nose.

4. Remove any remaining debris of the original

nose from the end of the piston rod.

5. Put the new piston nose in its recess in the

piston nose holding tool.

6. Turn the tool upside down and place the piston

assembly.

7. Slacken the piston button and remove it from

the piston rod. Slide the clamping tool down the

piston rod and press the 2 tools firmly against

each other. A slight rotation of the clamping tool

ensures a correct fitting.

8. Remove the tools away from the piston assembly.
9. Dip the piston nose in distilled water and slide

the piston into the syringe.

10. Hold the syringe upright with the threaded end

of the syringe resting on a flat, clean surface.

Slide the piston into the syringe until the piston

stops (the nose is level with the end of the syringe).

11. Slide the piston button against the body of the

syringe and tighten the Allen screw that secures

the piston button onto the piston rod.

12. Slide the piston up and down in the syringe five

or six times to ensure a smooth operation be-

tween the piston and syringe.

13. Make sure of the presence of the plastic cover

when re-mounting the syringe on the pump as

recommended below.

14. Remount the syringe on the pump as described

page 6-4.

Part Number Description

F4015064

5 x 500 µL piston
noses and
mounting tool kit

Piston rod

Clamping tool

Piston
nose

Piston nose holding tool