Installation, Assembling the probe – Gilson 402 User Manual

Page 27

Advertising
background image

3-3

Installation

3

2 Special adapters for the probe

1 D200 tips, pack of 10

1 D1000 tips, pack of 10

1 Transfer-tubing (outlet) non-tapered, FEP (2 mL)

1. Lay the tubing on the probe’s groove such that

the end with the small washer is just at the

orifice of the probe.

2. Push the tubing into the groove in the probe at

the 3 round access points using the rounded end

of a pencil (for example). Do not press on the

tubing with your fingernail.

3. Screw the tip-holder onto the end of the probe.
4. Plug the end of the control cable into the socket,

labeled ‘PROBE’ at the back of the 402.

5. Fit the clamp at the right-hand side of the 402

(see on the previous page).

6. Mount the probe on the clamp.
7. Connect the other end of the tubing to the outlet

of the valve (Single-syringe) or the Tee junction

(Dual-syringe).

8. Inform the 402 Software of the new configura-

tion using the Conf (Vol) menu.

9. Fit a tip to the holder. In use, when you change

a tip, wipe the end of the tip-holder first, to

remove any liquid from the orifice. Do not

twist the probe to the left, when fitting a tip,

otherwise the tip-holder will unscrew itself.

Assembling the Probe

(1) The small washer is important

in that it helps to ensure that

fluid does not leak from the

junction between the tip-holder

and the orifice of the probe.

(2) The special ‘S’ shaped metallic

clips, which are pre-mounted, are

for supporting the tubing on the

black control cable. The tubing

for use with the tip-holder will

not pass through the plastic

‘tunnels’ in the pre-mounted

cable supports. However, to

make the metallic clip easier to

attach, you can push these black

tubing-supports to the end of

the cable.