Operation - advanced use – Gilson 402 User Manual

Page 58

Advertising
background image

5-8

Operation - Advanced Use

5

When you specify left and right syringe capacities

(left must be > right), you must enter compatible

volumes. An error message is displayed if you try

to enter an incompatible volume, in which case

you may need to change the volume of the other

syringe, before entering the volume of the one

that provoked the error message.
• Correction Factor

To compensate for the effects of temperature and

viscosity, a factor may be applied to all liquid

volumes. This factor should be calculated from

gravimetric assays (see page 5-33).
With the Dual-syringe model, you must first

select the syringe to which the correction applies.

Starting from the Configuration Screen, press Syr,

then use Next (or Prev) to select the right-hand or

left-hand syringe.
Press Corr, you then see the Syringe Correction

Screen. Key in a factor between 0 and 9.9 (%), press

ENTER to confirm. Press ESC, then use Prev (or

Next) to select the other syringe (if fitted), enter

the correction in the same way; press ESC (twice)

to return to the Configuration Screen.

Flow Rates

Press the Flow soft-key; the Air Flow Screen appears.
Air Flow is the rate in ml/min. that air is drawn

into the transfer tubing when aspirating air.
Key in the required value, then press ENTER. Or,

press Deflt to use the default value (4 mL/min).
Press the ESC key to return to the Configuration

Screen or press Next to go to the Resv Flow Screen.
‘Resv asp:’ is the rate in mL/min. that solvent is

aspirated from the reservoir by the left syringe.
Key in the required value, then press ENTER. Or,

press Deflt to use the default value. Press the ESC

key to return to the Configuration Screen or press

Next to go to the Prime Flow Screen.
Prime Flow is the rate in mL/min. that liquid is

transferred by the syringe(s) when priming.
Key in the required value, then press ENTER. Or,

press Deflt to use the default value. Press the ESC

key to return to the Configuration Screen. The flow-

rate limits depend on the volume of the syringe(s)

and type of tubing used, see pages 5-31 and 5-32.

Software

When you turn the syringe

correction factor ON, a plus sign

(+) appears after the file number,

as a reminder.

Syringe Correction Screen

Air Flow Screen

Resv Flow Screen

Prime Flow Screen