Operation - advanced use, Entering the configuration (conf) – Gilson 402 User Manual

Page 56

Advertising
background image

5-6

Operation - Advanced Use

5

Software

4 Press the Next soft-key to go to the following

screen.

5 Press the Chge soft-key to choose an option or to

select from a list of possible values.

6 Press the ESC key to return to the level of the

previous screen or menu.

7 Press CLEAR to remove the value (before

pressing ENTER) and return to the previous

value.

8 Press Deflt to select the default value or option.

Entering the Configuration (CONF)

You have to tell the controlling software how your

402 Dilutor-Dispenser is configured (syringe sizes,

flow rates, etc.). Press the CONF key; you see the

Configuration Screen.

Volume(s)

Press the Vol soft-key, you see the Tubing Volume

Screen.
The use of the soft-keys is as follows:

Next is used to go to the next screen (from Tubing

Volume Screen),
Chge is used to select the tubing size,
Prev is used to go to the previous screen (from Air

Gap Volume Screen),

Check that the configured volume for the probe

tubing is the same as the size fitted. For the

standard probe, you can select from: 1, 2, 5, 10 mL

(tapered tubing).
When using disposable tips, you must select the size

of tip (Micro or Large Volume), and application

type for Micro Tips only (Air or Liquid). When you

use disposable tips, the ‘Extra Volume’ parameter is

automatically changed to ‘Air Push Volume’.

Press Key To Change

What Items

Vol

Volumes

Transfer Tubing, Air Gap.

Syr

Syringe

Sizes, Correction, and to change.

Flow

Flow-rates

Air, Prime, Reservoir.

Count

Counters

Number of cycles for syringe(s) and valve.

Etc

Switch On/Off

Beeper and Autorefill function.

Configuration Screen

Tubing Volume Screen