Maintenance, Cleaning the tee junction – Gilson 402 User Manual

Page 87

Advertising
background image

6-3

Maintenance

6

Cleaning

1) Pull the axle away from the valve main body.
2) Remove the ceramic stator from the valve axle.
3) Tap the valve inner body against a clean, solid

surface to remove the spring and PTFE spring cover.

4) Do not remove the inner ceramic surface from

the valve main body.

5) The parts can be cleaned with a non-abrasive cloth

dampened with alcohol, or by autoclaving them.

6) Dry the components using a lint-free cloth.
7) Check that the ceramic parts are dry and wait

for them to cool to room temperature before

reassembling the valve.

PTEF seal

Ceramic
stator

Valve axle

Spring

Ceramic and
PEEK rotor

PTFE-end piece

Valve axle guide

Cleaning the Tee Junction

The Tee junction is a single piece. Apart from dis-

mounting it from the 402, there is no disassembly

procedure. The Tee junction can be cleaned or

autoclaved.
1) Disconnect the inlet and outlet tubing from the

Tee junction.

2) Dismount the Tee junction and syringe assembly

and separate the syringe from the Tee junction.

3) Undo the two screws securing the Tee junction

to the 402 front plate and remove the Tee junc-

tion.

4) Autoclave the Tee junction.
5) Reinstall the Tee junction by following the

instructions given in Chapter 3.

Advertising