Operation - advanced use – Gilson 402 User Manual

Page 70

Advertising
background image

5-20

Operation - Advanced Use

5

For each dilution in the series:

Vt = Vs x Rd, Vd = Vt - Vs
For example, after selecting Dilution Ratio Mode, you

enter Sample Volume = 10 and Dilution Ratio = 3;

the software calculates the Total Volume as 30 and

the required Diluent Volume as 20.
1. When diluent comes from the reservoir (RESV);

aspirate the required volume of diluent (for all

dilutions, up to the capacity of the left-hand

syringe). If diluent comes from the probe, an air

gap is aspirated before the diluent, and the

required volume of diluent is aspirated (for all

dilutions, up to the capacity of the right-hand

syringe, after which the left-hand syringe is used).

2. Aspirate an air gap (specified in configuration).

3. Aspirate sample.

4. Dispense all into the recipient vessel.

5. Aspirate an air gap (specified in configuration).

6. Aspirate diluted sample.

7. Dispense all into the recipient vessel.

8. Aspirate an air gap (specified in configuration).

9. Aspirate diluted sample.

10. Dispense all into the recipient vessel.

11. And so on ...
Typical Task Entry
Key to press:
Key in the File # (e.g. 48),
Press ENTER
Press Edit

Press Dil
Press Serial

Press Chge to select source, press ENTER, then

Next.

Press Chge to select

mode, press ENTER,
Press Next, key in a

number of serial

dilutions, press ENTER.

Press Next, then key in the sample volume, press

ENTER.

Depending on the

mode, you see one of

three screens after

pressing Next.

Key in the value, press ENTER.

Software

Screens: