5-25

Operation - Advanced Use

5

Press ESC (twice).
Press ESC.

Pipette

You use this task to aspirate a series of fractions via

the probe, then dispense the total volume aspirated

into a recipient vessel. Each liquid aspirated is sepa-

rated by an air gap (if air gap is configured as > 0).
You may also aspirate an ‘Extra volume’ to ensure the

accuracy of the first fraction. This ‘Extra volume’,

which is aspirated with the first fraction, is dispensed

at the end of the process, into the waste.
1. Aspirate an air gap.
2. Aspirate the first fraction (with

‘Extra volume’, if specified).

3. Aspirate an air gap.
4. Aspirate the second fraction.
5. Aspirate an air gap.
6. Aspirate the third fraction.
7. Dispense all except ‘Extra volume’

into recipient vessel.

8. Dispense ‘Extra volume’ to waste.
Key to press
Key in the File # (e.g. 45), ENTER

Press Edit,
Press Pip, then select one of two modes, where:
(a) Fractions are equal,

(b) Fractions are different.
Press Chge to switch

modes. Press ENTER,

then Next.
Key in number of

fractions,

press ENTER, then

Next.

For SAME mode, key

in common volume,

for DIFFERENT

mode, key in volume

for fraction 1, press

Next, key in volume

for fraction 2, Next, etc.
Press ENTER, then Next.

Software

When you configure disposable

tips in air mode, ‘Extra volume’

is replaced by ‘Air push volume’

in this task.

1

air gap

2

fraction 1

3

air gap

4

fraction 2

5

air gap

6

fraction 3

7

result