Operation - advanced use – Gilson 402 User Manual

Page 68

Advertising
background image

5-18

Operation - Advanced Use

5

Dilute

Three types of dilution task are available ‘Normal’,

‘Unknown’, and ‘Serial’. Three modes are available

for ‘Normal’ and ‘Serial’: Total Volume, Diluent

Volume, and Dilution Ratio. The first action is to

select the type of dilution task that you are going to

use, by pressing the appropriate soft-keys.

Dilute Normal
You use Dilute Normal to dilute samples, with diluent

from the reservoir or via the probe, and add up to 12

reagents. There are 3 modes of entering the parameters,

in each mode you supply Sample Volume (Vs) plus

one of the following variables: Dilution Ratio (Rd),

Diluent Volume (Vd), and Total Volume (Vt). The

software calculates the remaining variables, taking

into account the volumes of the added reagents (Vr).

Vt = Vs x Rd, Vd = Vt - Vs - Vr

For example, after selecting Dilution Ratio Mode, you

enter Sample Volume = 10 and Dilution Ratio = 3,

Reagent Volume (total) = 9; the software calculates

the Total Volume as 30 and the required Diluent

Volume as 11. In any mode, pressing the Data soft-

key gives the calculated values for the other two

parameters.
1. Aspirate the required volume of diluent (in

opposite figure, from the reservoir).

2. Aspirate an air gap (specified in configuration).

3. Possibly, Aspirate reagent(s) then the sample

(with an air gap between each liquid).

4. Dispense all into recipient vessel.

Typical Task Entry
Key to press:
Key in the File # (e.g. 48),
Press ENTER
Press Edit
Press Dil
Press Normal
Press Chge to select

source, press ENTER,

then Next.
Press Chge to select the

mode, press ENTER,

Software

diluent

air gap

sample

1

4

2

3

result

Screens:

If you specify a volume of sample

greater than that of the right-

hand syringe, the left-hand will

be used to aspirate the sample.