Technical data appendix, Syringes performance specifications, Protocol – Gilson 402 User Manual

Page 111: Hydraulics

Advertising
background image

Appendix C-3

Technical Data

Appendix

C

Hydraulics

Syringes

Performance Specifications

The following values are obtained using a defined

protocol. If this protocol is followed, the obtained

values must be inferior or equal to the listed figures.

Syringe capacity (µL)

100

250

500

1000

5000

10000

25000

Maximum Flow (mL/min)

4

10

20

40

120

120

120

Minimum Specified Volume (µL)

10

25

50

100

500

1000

2500

Nominal Volume

(µL)

100

250

500

1000

5000

10000

25000

Protocol

Configuration: 1 mL tubing for volumes up to 1 mL,

tapered transfer-tubing of 2 mL (standard accessory)

for volumes up to 2 mL, 5 mL or 10 mL tapered

transfer tubings are used for volumes up to 10 mL.

Transfer of de-ionized water using the Pipette Task,

aspiration of liquid from a vial and delivery at the

same flow rate into another vial containing water.

Syringe

Volume Systematic error Random error

capacity (µL) pipetted (µL)

%

µL

%

µL

10

1.8

0.18

0.8

0.08

100

50

1.5

0.75

0.25

0.125

100

1

1

0.15

0.15

20

1.5

0.38

0.8

0.2

250

125

1.0

1.25

0.25

0.32

250

0.8

2

0.15

0.38

50

1.5

0.75

0.50

0.25

500

250

1.0

2.5

0.25

0.63

500

0.8

4

0.15

0.75

100

1.5

1.5

0.50

0.5

1000

500

0.8

4

0.25

1.3

1000

0.6

6

0.15

1.5

500

1.2

6

0.5

2.5

5000

2500

0.8

20

0.25

6.3

5000

0.6

30

0.15

7.5

1000

1.2

12

0.5

5

10000

5000

0.8

40

0.25

12.5

10000

0.5

50

0.15

15

2000

1.0

25

0.50

5

25000

12500*

0.8

100

0.25

31.3

25000*

0.5

125

0.15

37.5

* Dispensing mode only.

The values are obtained by

pipetting de-ionized water

(grade 3, ISO3696) at 21 °C ± 0.5 °C,

relative humidity > 50 % and

athmospheric pressure 1013 MPa.