Maintenance

6

Cleaning

Cleaning the Syringe Assembly

You must dismount the syringe and piston assembly

from the instrument and then separate the syringe and

piston assembly before cleaning or autoclaving them.
1) Dismount the Syringe assembly from the instrument.
2) Undo the piston securing screw to free the

piston assembly from the piston operating rod.

3) Lower the piston operating rod; for Gilson soft-ware

there is a change syringe program that lowers the

piston operating rod - use this software option to

lower the piston operating rod away from the

syringe piston. Refer to your software User’s

Guide for the precise sequence. This allows you to

dismount and mount the syringe and valve/Tee

assembly without the risk of damaging the syringe.

4) Undo the syringe from the body of the valve or Tee.
5) Remove the piston and syringe assembly from

the pump.

6) Pull the piston out of the syringe body. For large

syringes; 5, 10 and 25 mL syringes, you must

screw the piston securing screw back into the

bottom of the piston and pull on the screw to

remove the piston from the syringe barrel.

7) Clean or autoclave the two parts (except the 25 mL

syringe).

8) Ensure that the syringe and piston assembly

have cooled to room temperature before reas-

sembling the syringe and piston. Ensure that the

piston and the piston nose are free from dust or

contamination before introducing the piston into

the syringe. Wet the piston nose (water will

suffice) before inserting the piston in the syringe.

Insert the piston nose in the syringe body and

push the piston into the syringe.

9) Remount the syringe assembly on the valve or Tee.

Screw the threaded end of the syringe body into

the valve or Tee.

10) Lower the piston so that it comes into contact

with the piston holding screw.

11)Tighten the piston holding screw so that the

piston assembly is secured against the piston

operating rod.

The 25 mL syringe is not

autoclavable, all of the other

size syringes are autoclavable.

Advertising