MTS Series 793 User Manual
Page 135
Advertising
Displacement Sensor Calibration
MTS Series 793 Tuning and Calibration
Calibration Procedures
135
Conditioner calibration
values
For LVDT calibration, set the following initial conditioner calibration
values on the Calibration submenu or tab:
C
ONTROL
S
ETTING
Polarity
Normal
Pre-Amp Gain
1.0
Post-Amp Gain
1.5
Excitation
10 volts
Phase
45º
Delta K
(Gain/Delta K calibration only)
1
Fine Zero
0