Task 8 establish the shunt calibration reference – MTS Series 793 User Manual

Page 182

Advertising
background image

MTS Series 793 Tuning and Calibration

Force Sensor Calibration

Calibration Procedures

182

A. Enter the full scale force values in the Fullscale Min/Max entry boxes.

B. Adjust Neg Tension for the required tension sensitivity value (+mV/

V).

C. Adjust Pos Compression for the required compression sensitivity

value (-mV/V).

D. Adjust Excitation for the required calibration excitation value (DC).

Task 8

Establish the shunt calibration reference

Each resistive bridge type transducer (DC sensor) uses a shunt resistor to check
the calibration accuracy of the sensor/conditioner combination. Each DC
conditioner supports a shunt resistor.

1. Turn off hydraulic power.

2. Remove the load standard.

3. Turn on hydraulic power.

4. Zero the force sensor output.

Standalone FlexTest SE Controllers:

Adjust the Setpoint for a 0 kN output. The sensor output must be 0.000 kN
for a proper shunt calibration. If not, return to

“Set the zero and offset”

on

page 163.

Automated Controllers:

Adjust the Manual Cmd slider on the Manual Command window for a 0
kN output. The sensor output must be 0.000 kN for a proper shunt
calibration. If not, return to

“Set the zero and offset”

on page 163.

5. Change the control mode to displacement.

Standalone FlexTest SE Controllers: Select Status > Control Mode >
Disp.

Automated Controllers: On the Manual Command window, change
Control Mode to Displacement.

Shunt calibration cannot be performed on a sensor when it is in control of
the servo loop.

Advertising