Task 9 save the calibration – MTS Series 793 User Manual

Page 186

Advertising
background image

MTS Series 793 Tuning and Calibration

Force Sensor Calibration

Calibration Procedures

186

B. Ensure that Shunt State (+) is set to On. Polarity is always positive.

Note the Shunt Ref (+) value, it should be 60%–90% (80% is ideal) of
the calibrated range of the sensor.

C. Click Update to copy the current shunt reference value into the Shunt

Reference box.

D. Set Shunt State (+) to Off.

Automated Controllers:

A. In Station Setup select the appropriate force channel on the navigation

panel., click the Channel Input Signals icon, and then click the Shunt
tab.

B. Select the shunt type. Use (+) polarity if you are not sure what to select.

C. In the Current Shunt Value box, click the On button. Note the

Current Shunt Value, it should be 60%–90% (80% is ideal) of the
calibrated range of the sensor.

D. Click Update to copy the Current Shunt Value into the Shunt

Reference Value box, and then click Off.

Task 9

Save the calibration

It is important that you save your sensor calibration values.

Standalone FlexTest SE Controllers: Select Setup > Open/Save Parameters >
<<Save>>

Automated Controllers: On the Station Setup window Inputs panel, click the
Calibration tab, and then Save. This saves current calibration values on the
Calibration, Sensor, and Shunt tabs to the sensor calibration file.

Note

Systems equipped with full range conditioners (e.g., Model 493.25
DUCs) do not require or support multiple ranges. They use one range
that is typically set to 100%.

Advertising