Displacement Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

139

Before beginning, ensure the following are true:

The Fine Zero adjustment is set to 0.

Standalone FlexTest SE Controllers: Select Setup > Disp > Calibration
> Fine Zero

Automated Controllers: Adjust Fine Zero on the Calibration tab.

The Manual Offset is set to 0.

Standalone FlexTest SE Controllers: Select Setup > Disp > Offset/Zero >
Manual Offset

Automated Controllers: Adjust Manual Offset on the Offset/Zero tab.

The Electrical Zero adjustment is set to 0.

Standalone FlexTest SE Controllers: Select Setup > Disp > Offset/Zero >
Electrical Zero

Automated Controllers: Adjust Electrical Zero on the Offset/Zero tab