Displacement Sensor Calibration
MTS Series 793 Tuning and Calibration
Calibration Procedures
139
Before beginning, ensure the following are true:
•
The Fine Zero adjustment is set to 0.
Standalone FlexTest SE Controllers: Select Setup > Disp > Calibration
> Fine Zero
Automated Controllers: Adjust Fine Zero on the Calibration tab.
•
The Manual Offset is set to 0.
Standalone FlexTest SE Controllers: Select Setup > Disp > Offset/Zero >
Manual Offset
Automated Controllers: Adjust Manual Offset on the Offset/Zero tab.
•
The Electrical Zero adjustment is set to 0.
Standalone FlexTest SE Controllers: Select Setup > Disp > Offset/Zero >
Electrical Zero
Automated Controllers: Adjust Electrical Zero on the Offset/Zero tab