MTS Series 793 User Manual
Page 184
MTS Series 793 Tuning and Calibration
Force Sensor Calibration
Calibration Procedures
184
8. FlexTest SE/LE, FlexTest GT, TestStar IIm Controllers only: If you do not
have transducer ID modules on your sensor cables, install the shunt
calibration resistor as follows:
A. Select the appropriate shunt calibration resistor.
B. Bend the resistor leads 90º for a 0.3 inch separation.
C. Cut the resistor leads 0.12 inch from the bend.
D. Insert the resistor into the connector solder cups and solder.
E.
Complete and attach a shunt calibration label as specified on the
493.40/41 Carrier I/O Shunt Calibration Kit (MTS PN 100-028-185).
F.
Install the shunt cal resistor/connector assembly into the appropriate
slot of the SHUNT CAL connector on the front panel of the
appropriate I/O Carrier Module.