MTS Series 793 Tuning and Calibration

Force Sensor Calibration

Calibration Procedures

184

8. FlexTest SE/LE, FlexTest GT, TestStar IIm Controllers only: If you do not

have transducer ID modules on your sensor cables, install the shunt
calibration resistor as follows:

A. Select the appropriate shunt calibration resistor.

B. Bend the resistor leads 90º for a 0.3 inch separation.

C. Cut the resistor leads 0.12 inch from the bend.

D. Insert the resistor into the connector solder cups and solder.

E.

Complete and attach a shunt calibration label as specified on the
493.40/41 Carrier I/O Shunt Calibration Kit (MTS PN 100-028-185).

F.

Install the shunt cal resistor/connector assembly into the appropriate
slot of the SHUNT CAL connector on the front panel of the
appropriate I/O Carrier Module.