Force Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

173

A. Use the Dial to adjust the Setpoint for a 0% command.

B. Record the force standard’s readout value in the 0% row of your record

sheet.

C. Enter the corresponding conditioner feedback reading.

D. Use the Dial to adjust the Setpoint for a -2% tensile force command.

E.

Record the force standard’s readout value and its corresponding
conditioner feedback reading in the -2% row of your record sheet.

F.

Repeat steps D and E for other tensile force commands (typically at -4,
-6, -8, -10, -20, -40, -70, and -100 percent of full scale).

Automated Controllers:

A. Adjust the Manual Cmd slider for a 0% command.

B. Record the force standard’s readout value and corresponding

conditioner feedback reading at 0% command.

C. Adjust the Manual Cmd slider for a -2% tensile force command.

D. Record the force standard’s readout value and its corresponding

conditioner feedback reading in the -2% row of your record sheet.

E.

Repeat steps E-G for other tensile force commands (typically at -4, -6, -
8, -10, -20, -40, -70, and -100 percent of full scale).

7. Record conditioner feedback readings at predetermined compressive force

command points.

Standalone FlexTest SE Controllers:

A. Use the Dial to adjust the Setpoint for a +2% compressive force

command.

B. Record the force standard’s readout value in the +2% row of your

record sheet.

C. Enter the corresponding conditioner feedback reading.

D. Repeat steps A-C for other compressive force commands (typically at

+4, +6, +8, +10, +20, +40, +70, and +100 percent of full scale).

Automated Controllers:

A. Adjust the Manual Cmd slider for a +2% compressive force

command.