MTS Series 793 User Manual

Page 157

Advertising
background image

Force Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

157

Range support

Full-range conditioners (e.g. Model 493.25 DUCs) do not require individual
ranges.

Ranged conditioners (e.g. Model 493.21/21B DUCs and Model 497.22 DC
Conditioners) support up to four ranges for force sensor calibration. Force can be
calibrated to any range within the full-scale capacity of the sensor. Also, ranges
do not have to be centered on zero.

What you will need

You will need the following items to calibrate a a force sensor.

A load standard can be a calibrated force sensor with its own electronics or
it can be a calibrated set of dead weights.

A DVM to monitor the output of the load standard.

Note

All calibration tools should be calibrated to an industry standard.

Note

This calibration procedure calibrates the DC conditioner for a force
sensor of ±10 kN. You will need to adjust the procedure to accommodate
your force sensor.