Force Sensor Calibration
MTS Series 793 Tuning and Calibration
Calibration Procedures
161
Note
Sensor calibration and range information can be edited on the Sensor
tab, located on the Station Setup window Inputs panel.
See “How to Create a Sensor File” in Chapter 2: Station Manager of the
MTS Series 793 Control Software manual.
Conditioner calibration
values
For force sensor calibration, set the following initial conditioner calibration
values.
C
ONTROL
S
ETTING
Polarity
Normal
Pre-amp Gain
250
480 (for Model 493.25 conditioner)
Post-amp Gain
2
1 (for Model 493.25 conditioner)
Total Gain
500
Excitation
10 volts
Fine Zero
0
Zero/Balance
0
Delta K (Gain/Delta K calibration
only)
1