Force Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

163

Task 4

Set the zero and offset

This task records the load standard readout as the zero reference.

Using a load standard

Standalone FlexTest SE Controllers: Adjust the Setpoint for 0 kN. Then zero
the load standard readout. Select Status > Setpoint

Automated Controllers: Adjust the Manual Cmd slider on the Manual
Command window
for 0 kN. Then zero the load standard readout.

Using dead weights

Standalone FlexTest SE Controllers: Remove all dead weights, and then click
Auto Offset on the Offset/Zero menu. Select Setup > Force > Offset/Zero >
<<Auto Offset>>

Automated Controllers: Remove all dead weights, and then click Auto Offset
on the Offset/Zero tab (Inputs panel).

Task 5

Gain/Delta K Calibration

If you using Gain/Delta-K for your calibration type, complete the following
procedure. If not, complete

“Gain/Linearization Calibration” on page 170

or

“Millivolt/Volt Calibration”

on page 179.

Calibrate tension

Force sensors can be calibrated so that a positive output represents actuator
compression and a negative output represents actuator tension, and vice versa.

You calibrate the negative side of the output with gain and the positive side of the
output with Delta K.