MTS Series 793 Tuning and Calibration

How to Perform Advanced Tuning Techniques

Tuning Procedures

90

G.

For Upper Limit and Lower Limit, set the upper and lower limits for
actuator travel during auto-tuning.

Basic auto-tuning exercises within 80% of these limits.

Advanced and Advanced Only auto-tuning functions exercise the
actuator within 20% of these limits. If the limits are exceeded, the auto-
tuner will quit and trip an interlock.

H. For Advanced and Advanced Only auto-tuning—Set the Tracking%

value.

The 50% default setting is appropriate for most systems.

I.

For Advanced and Advanced Only auto-tuning—Set the Sweep Freq.

The Sweep Freq sets the upper frequency limit of the sine sweep. The
20 Hz default setting is a good starting point.

Pressing the Run button will put the actuator in motion.

A moving actuator can injure anyone in its path.

Always clear the actuator area before pressing the Run button.

Note

While auto-tuning is in process you cannot changes channels or control
modes. Also, the active and auto-tuning signal’s tuning parameters will
be disabled.

5. Press Run to start auto-tuning, and if desired, accept tuning values.

While auto-tuning is in progress, the controller displays:

“Auto Tuning in Progress...To terminate auto-tuning before
completion, press <Stop>.”

When auto-tuning is complete, the controller displays:

“Auto Tuning for this mode has completed successfully.”

Also, when auto-tuning is complete, the results are displayed showing
Current tuning values and Reference values.

Press Accept to apply the Reference values.

WARNING