Displacement Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

149

F.

Repeat steps D and E for other retraction commands (typically at -4, -6,
-8, -10, -20, -40, -70, and -100 percent of full scale).

Automated Controllers:

A. Adjust the Manual Cmd slider for a 0% command.

B. Record the dial indicator value for the 0% command.

C. Enter the corresponding conditioner feedback reading in the

Conditioner column at the appropriate row in the window.

D. Adjust the Manual Cmd slider for a -2% retraction command.

E.

Record dial indicator value and conditioner feedback at the -2% row of
your record sheet.

F.

Repeat steps D and E for other retraction commands (typically at -4, -6,
-8, -10, -20, -40, -70, and -100 percent of full scale).

5. Record dial indicator and conditioner feedback readings at predetermined

extension command points.

Standalone FlexTest SE Controllers:

A. Use the Dial to adjust the Setpoint for a +2% extension command.

B. Record the dial indicator and conditioner feedback values at the +2%

row of your record sheet.

C. Repeat steps A and B for other extension commands (typically at +4,

+6, +8, +10, +20, +40, +70, and +100 percent of full scale).

Automated Controllers:

A. Adjust the Manual Cmd slider for a +2% extension command.

B. Record the dial indicator value and conditioner feedback at the +2%

row of your record sheet.

C. Repeat steps A and B for other extension commands (typically at +4,

+6, +8, +10, +20, +40, +70, and +100 percent of full scale)

6. Turn off system hydraulics.