Task 7 save the calibration settings – MTS Series 793 User Manual

Page 212

Advertising
background image

MTS Series 793 Tuning and Calibration

Extensometer Calibration

Calibration Procedures

212

5. Perform shunt calibration.

Standalone FlexTest SE Controllers:

Select Setup > Strain > Calibration > Select Cal Type >

A. Select required Shunt Display units.

B. Ensure that Shunt State (+) is set to On. Polarity is always positive.

Note the Shunt Reference (+) value, it should be 60%–90% (80% is
ideal) of the calibrated range of the sensor.

C. Click Update to copy the current shunt reference value into the Shunt

Reference box.

D. Set Shunt State (+) to Off.

Automated Controllers:

A. In Station Setup select the appropriate strain channel on the navigation

panel., click the Channel Input Signals icon, and then click the Shunt
tab.

B. Select the shunt type. Use (+) polarity if you are not sure what to select.

C. In the Current Shunt Value box, click the On button. Note the

Current Shunt Value, it should be 60%–90% (80% is ideal) of the
calibrated range of the sensor.

D. Click Update to copy the Current Shunt Value into the Shunt

Reference Value box, and then click Off.

Task 7

Save the calibration settings

It is important that you save your sensor calibration values.

Standalone FlexTest SE Controllers: Select Setup > Open/Save Parameters >
<<Save>>

Automated Controllers: Click Update File on the Calibration tab

This saves the current calibration values to your current sensor calibration file.

Advertising