Task 2 create a calibration file, Task 3 assign a calibration file, Task 4 turn on hydraulic pressure – MTS Series 793 User Manual

Page 192

Advertising
background image

MTS Series 793 Tuning and Calibration

Encoder and Temposonics Calibration

Calibration Procedures

192

Task 2

Create a calibration file

This task creates a sensor calibration file.

Note

Encoder and Temposonics sensors do not require ranges, their
resolution is constant.

See “How to Create a Sensor File” in the MTS Series 793 Control Software
manual.

Setting device resolution

In the Resolution box, enter the resolution supplied in the device’s
documentation.

Note

If you want to use units different than those supplied in the sensor
documentation, enter the full scale and resolution in the supplied units
first, and then switch to the desired full-scale units. The units conversion
will be calculated automatically.

Task 3

Assign a calibration file

This task links a sensor calibration file to a hardware resource. The purpose for
this is to select one of the sensor ranges for the input signal definition.

See “How to Assign a Sensor File” in the MTS Series 793 Control Software
manual.

Task 4

Turn on hydraulic pressure

This task sets up the Control Panel so you can turn on the hydraulic pressure.

Do not place any part of your body in the path of a moving actuator.

A crush zone exists between the actuator and any equipment in the path of
its movement. Immediate and unexpected actuator response is possible
when you apply hydraulic pressure to your system.

Stay clear of the actuators when applying hydraulic pressure.

See

“Turn on hydraulic pressure”

on page 136 for a detailed procedure.

WARNING