How to balance dual valves, How to balance dual valves 63 – MTS Series 793 User Manual

Page 63

Advertising
background image

How to Balance the Servovalve

MTS Series 793 Tuning and Calibration

Tuning Procedures

63

A. In the Station Setup window, click

.

Three-stage valves: In the Drive panel, click the Valve tab.

B. While observing Displacement Abs. Error on the Meters panel, adjust

the Valve Balance control until Displacement Abs. Error is zero.

6. Return displacement I Gain to its original setting.

Standalone FlexTest SE Controllers: Select Tuning > Manual > I Gain

Automated Controllers: I

A. In the Station Setup window, click

.

B. In the Tuning panel, click the Adjustments tab.

How to Balance Dual Valves

Use the following procedure if you need to balance dual servovalves.

1. Get things ready.

A. If necessary, select the tuning access level.

B. Remove any specimen.

C. Determine which servovalve of the dual valve pair you want to balance

first, then mount a blocking plate on the port of the other servovalve.

Note

When installing the blocking plate ensure that its holes are aligned with
the servovalve port holes and a gasket is used.

D. Apply hydraulic pressure.

Selecting the Enable Manual Command allows you to manually position
actuators.

A moving actuator can injure anyone in its path.

Always clear the actuator area before selecting Enable Manual Command.

WARNING

Advertising