MTS Series 793 User Manual

Page 203

Advertising
background image

Extensometer Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

203

4. Set the calibrator micrometer head for 80% of the negative full-scale value.

5. Note the monitored strain signal value.

6. If the signal value does not match the set value, adjust the transducer

excitation voltage to achieve an 80% value (–8 V DC). Record your final
strain signal value for the 80% output.

Note

After determining an approximate excitation value, round it to convenient
value (e.g., 8.00 V DC). Use Post-amp Gain to make fine adjustments of
calibration.

Calibrate the positive

output (compression)

To calibrate the positive extensometer output:

1. Adjust the calibrator between zero and 100% of the extensometer’s full-

scale range three times. This exercises the extensometer to remove any
hysteresis.

Note

Before clicking Auto Zero, ensure that any offsets due to Auto Offset
and Manual Offset actions are eliminated

2. Use Auto Zero to zero transducer output.

Standalone FlexTest SE Controllers: Select Setup > Strain > Offset/Zero
> <<Auto Zero>>

Automated Controllers: Click Auto Zero on the Calibration tab to set
transducer’s amplified output to 0.000 V DC.

3. Set the calibrator micrometer to 80% of the positive full scale value.

4. Note the monitored strain signal value.

5. Adjust Delta K to achieve an 80% value (+8 V DC).

Note

The Delta K value can be adjusted above or below 1.000. This means it
will correct for an error above or below 8 volts.

Compare recorded

data points to

calibration sheet

Your sensor should include a calibration data sheet that shows the data point
tolerance.

1. Record the negative and positive outputs values at the 20%, 40%, 60%, and

100% data points.

2. Compare your recorded output values to the calibration data sheet that

accompanied your sensor.

3. Make sure your current values fall within the permissible variation.

Advertising