MTS Series 793 User Manual

Page 101

Advertising
background image

How to Perform Advanced Tuning Techniques

MTS Series 793 Tuning and Calibration

Tuning Procedures

101

3. In Station Setup, click the Channel Drive icon to display the Drive panel,

and then click the Conditioner tab.

4. From the Service Port Output, select one of signals described in the

following Service Port Output Signals list.

Note

The Spool Position signal is available to the digital meters or software
scope.

Monitoring inner-loop

signals on FlexTest IIm

Controllers

Inner-loop signals can be monitored with external meters via the tip jacks on the
Model 497.15 3-Stage Valve Driver module. You must specify which signal is
available from the tip jacks with the Monitor Mux Output list.

Service Port Output Signals

S

IGNAL

D

ESCRIPTION

Valve Current

The final output to the servovalve. The

voltage represents the current output
[10 V (DC) = 50 mA; standard].

Demod Filter Output

The analog output of the demodulator

prior to gain.

Innerloop Command

The input signal to the 3-Stage Valve

Driver (the test program command).

Spool Position

The Conditioner Out signal summed
with the Spool Zero signal.

Spool Offset

Signal that is summed with spool position

to remove any DC offset. ±10 V
represents ±4 V of zero summing.

Preamp Output

The raw AC input from the spool LVDT.

Conditioner Out

The conditioned feedback signal from the
servovalve LVDT.

Voltage Reference (5V)

Internal board reference test only.