MTS Series 793 Tuning and Calibration

Extensometer Calibration

Calibration Procedures

202

5. Use Auto Zero to zero transducer output.

Standalone FlexTest SE Controllers:

Select Setup > Strain > Offset/Zero > <<Auto Zero>>

Automated Controllers:

Click Auto Zero on the Calibration tab to set transducer’s amplified output
to 0.000 V DC.

Calibrate the negative

output (tension)

This task calibrates the extensometer negative output using the calibration Gain
controls. Since the Gain setting will affect your Delta K setting, you should
always calibrate the negative side first.

To calibrate the negative extensometer output:

1. Set an initial nominal Gain setting of 500.

2. Adjust the calibrator between zero and -100% of the extensometer’s full-

scale range three times. This exercises the extensometer to remove any
hysteresis.

3. Use Auto Zero to zero transducer output.

Standalone FlexTest SE Controllers: Select Setup > Strain > Offset/Zero
> <<Auto Zero>>

Automated Controllers: Click Auto Zero on the Calibration tab to set
transducer’s amplified output to 0.000 V DC.

mechanical input

Indicated output

Δ

K > 1

Δ

K = 1

Δ

K < 1

Gain