Extensometer calibration: abbreviated procedure – MTS Series 793 User Manual
Page 197
Advertising
Extensometer Calibration
MTS Series 793 Tuning and Calibration
Calibration Procedures
197
Extensometer Calibration: Abbreviated Procedure
Task 1, “Get things ready,”
on page 198.
Task 2, “Create a sensor calibration file,”
on page 199
Task 3, “Adjust offset,”
on page 200.
Task 4, “Gain/Delta K Calibration,”
Note
The Gain/Linearization Calibration procedure applies only to controllers
equipped with full-range conditioners (e.g., Model 493.25 DUC module).
Task 5, “Gain/Linearization Calibration,”
Task 6, “Establish the shunt calibration reference,”
on page 210.
Task 7, “Save the calibration settings,”
on page 212.