MTS Series 793 User Manual

Page 143

Advertising
background image

Displacement Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

143

The Total Gain value is calculated by multiplying the Pre-Amp and Post-Amp
gain values. If the total desired gain amount is known (from a calibration sheet),
you can enter the amount in the Total Gain box and the software will calculate
the Pre-Amp and Post-Amp gain values automatically.

When you are calibrating an AC conditioner, use the Post-Amp gain control to
increase gain. If more gain is needed, you must disable hydraulics and increase
the Pre-Amp gain. You can then turn on hydraulics and continue to adjust the
Post-Amp gain.

Note

This procedure assumes a positive command extends the actuator and a
negative command retracts the actuator. If not, use the opposite polarity
for any values given.

Procedure

1. Select Gain/Delta-K for Cal Type.

Standalone FlexTest SE Controllers: Select Setup > Disp > Calibration
> Cal Type > Gain/Delta-K

Automated Controllers: Select Gain/Delta-K for Cal Type on the
Calibration tab of the Inputs panel.

2. Apply a retraction command that is 80% of the full scale range.

Standalone FlexTest SE Controllers: Standalone FlexTest SE
Controllers:

A. Enable the front panel Dial.

B. Use the Dial to adjust the Setpoint for 80% of the full scale range.

C. Verify that your LVDT displacement signal equals 80% of the full scale

range.

Automated Controllers:

A. Open the Manual Command window, and then click on the Enable

Manual Command.

B. Adjust the Manual Cmd slider on the Manual Command window for

80% of the full scale range.

C. Use the Station Signals panel to verify that your LVDT displacement

signal equals 80% of the full scale range.

During the initial calibration and tuning of your system, it may take repeated
adjustment for the two values to match.