Displacement Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

147

Note

The Gain/Linearization Calibration procedure that follows applies only to
controllers equipped with full-range conditioners (e.g., Model 493.25
DUC module).

Task 7

Gain/Linearization Calibration

If you are using Gain/Linearization for your calibration type, complete the
following procedure. If not, complete Task 6 Gain/Delta K Calibration on
page 142
.

Important

Using linearization data requires specific conditioner zeroing
practices. Ensure that Electrical Zero
is set to zero and the
Electrical Zero Lock box is set (enabled). Adjusting electrical zero
after calibration may invalidate linearization data.

Initial LVDT calibration

For initial calibration of an LVDT complete the following procedure:

1. Select Gain/Linearization for Cal Type.

Standalone FlexTest SE Controllers: Select Setup > Disp. > Calibration
> Cal Type > Gain/Linearization

Automated Controllers: Select Gain/Linearization for Cal Type on the
Calibration tab of the Inputs panel.

2. Apply a retraction command that is 80% of the full scale range.

Standalone FlexTest SE Controllers:

A. Enable the front panel Dial.

B. Use the Dial to adjust the Setpoint for 80% of the full scale range.

C. Verify that your LVDT displacement signal equals 80% of the full scale

range.

Automated Controllers:

A. Open the Manual Command window, and then click on the Enable

Manual Command.

B. Adjust the Manual Cmd slider on the Manual Command window for

80% of the full scale range.

C. Use the Station Signals panel to verify that your LVDT displacement

signal equals 80% of the full scale range. During the initial calibration
and tuning of your system, it may take repeated adjustment for the two
values to match.