Import hwi defaults, System options settings, Vme bus board settings – MTS Series 793 Utility Software User Manual

Page 38

Advertising
background image

Import HWI Defaults

Previously exported Hwi Defaults files can be imported. After HwiDefaults are imported, the Hwi
Editor application will use those default values for any subsequent Hwi files that you create.

Note:

Hwi Default files use a “.hwidef” file extension.

Use one of the following methods to import a previously exported Hwi Defaults file.

On the File menu, click Import HWI Defaults.

OR

Right-click the HWI Defaults icon and click Import HWI Defaults.

OR

Right-click the Controller icon and click Import HWI Defaults.

System Options Settings

System Options icon

Velocity Limiter: When enabled (true), limits actuator movement to a maximum of 10 mm/sec.
when displacement is commanded from the RSC or handset. This feature provides compliance
to CE requirements. When Velocity Limiter=True, the manual command control is locked.

Interlocks: When the Hwi File Editor reads existing .hwi files, this value represents the number of
interlock chains in the file. If you are creating a new .hwi file, enter the desired number of interlock
chains.

VME Bus Board Settings

FlexTest
60/100/200

(494.06/.10/.20)

FlexTest
40
(494.04)

FlexTest
SE
(493.02)

Aero ST
(493.20)

FlexTest
GT,
Teststar
IIM
(493.10)

FlexTest
IIM
(498.xx)

VME Board

X

X

X

X

X

X

Processor

24 MTS Series 793 Utility Software