Description

Item

(Should this be read only?)Type: Solenoid, On/Off Solenoid,
Proportional, Remote.

If you select Proportional, you configure the proportional output with
the controller software. The output signal can be ramped from 20
mA (minimum) to 700 mA (maximum) which corresponds with 50
psi (0.4 MPa) and 3000 psi (21 MPa). By default, low pressure is
factory set at 750 psi (5.25 MPa) and high pressure is set at 3000
psi (21 MPa).

If you select On/Off Solenoid, only two HSM control buttons (Off/On)
will appear on the Station Manager Station Controls panel. For the
other types, three HSM control buttons will appear (Off/Low/High).

Remote: select if more than one controller share the same HSM and
the control of the HSM is performed remotely on another controller.

Low Percent: 25. HSM low pressure setting as a percentage of
full-scale.

High Percent: 100. HSM high pressure setting as a percentage of
full-scale.

HSM Rate: Slow = 4 seconds. Fast = 2 seconds.

Time to achieve low pressure from zero pressure = (Low Percent
value) x (HSM Rate value).

Time to achieve high pressure from low pressure = (High Percent
value x HSM Rate value) - (Time to achieve low pressure).

Example: Suppose Low Percent value = 25, High Percent value =
100, and HSM Rate = Slow (4 seconds).

Time to achieve low pressure = 0.25 x 4 = 1 second.

Time to achieve high pressure = (1 x 4) - (1) = 3 seconds.

Connector: J43,J29.

Interlock Interface icon

Assign to Interlock: Set to 1.

Name: Describes this resource as it will appear in Station Builder
resource lists when opened against an .hwi file.

Analog Input/Output icons

Changing this name may invalidate existing station configurations.

For FlexTest SE Controllers, the first two analog output resources
are displayed as “Monitor I” and “Monitor 2” by default. These
resources represent BNC connectors on the Controller’s front panel.
The third analog output resource (Anlg Out) is the BNC connector
on the system board.

Address: Not applicable.

MTS Series 793 Utility Software 43