MTS Series 793 Utility Software User Manual

Page 49

Advertising
background image

Description

Item

panel. For the other types, three HSM control buttons will appear
(Off/Low/High).

If you select Proportional, you configure the proportional output
with the controller software. The output signal can be ramped
from 20 mA (minimum) to 700 mA (maximum) which
corresponds with 50 psi (0.4 MPa) and 3000 psi (21 MPa). By
default, low pressure is factory set at 750 psi (5.25 MPa) and
high pressure is set at 3000 psi (21 MPa).

Low Percent: 25. HSM low pressure setting as a percentage of
full-scale.

High Percent: 100. HSM high pressure setting as a percentage
of full-scale.

HSM Rate: Slow = 4 seconds. Fast = 2 seconds.

Time to achieve low pressure from zero pressure = (Low Percent
value) x (HSM Rate value).

Time to achieve high pressure from low pressure = (High
Percent value x HSM Rate value) - (Time to achieve low
pressure).

Example: If Low Percent value = 25, High Percent value = 100,
and HSM Rate = Slow (4 seconds):

Time to achieve low pressure = 0.25 x 4 = 1 second.

Time to achieve high pressure = (1 x 4) - (1) = 3 seconds.

Connector: J43,J29.

Interlock Interface icon

Assign to Interlock: Set to 1.

Gate Interlock: True or False, True = the controller responds to
the gate interlock signal on pins 3 and 4 of J29. (Default=True.)
False=J29 pins 3 and 4 are reserved for Program Stop interlock.

CAUTION:

Failure to enable the gate interlock option on
systems that can operate in Slow (low-flow) mode
can result in unexpected actuator movement when
switching the system from Slow mode to Fast mode.

Unexpected actuator movement can result in injury
to personnel or damage to the equipment.

Enable the gate interlock option for systems that
include a Slow (low-flow) mode for specimen
installation.

MTS Series 793 Utility Software 35