Miscellaneous device settings, Temperature controller settings – MTS Series 793 Utility Software User Manual

Page 92

Advertising
background image

Miscellaneous Device Settings

Use the drop-down lists in the Miscellaneous tab to add various external devices to the .hwi file.
The devices listed here are not automatically added by the Hardware Detect option.

Temperature Controller Settings

Description

Item

This indicates a Eurotherm Series 2200 or Series 2400 temperature
controller is connected to the controller. A Model 409 Temperature
controller may contain multiple Eurotherm controller channels.

Temperature Control icon

A GRESIII board or Model 494.47 Dual UART/Encoder card is
required to support a temperature controller.

Type: Enter the name of the temperature controller as desired.

Interlock Temperature (deg C) Define the default temperature. This
temperature will be sent to the temperature controller whenever the
station is interlocked or not loaded.

Use Interlock Temperature: Select True to use the interlock
temperature feature.

Com Timeout Delay (sec): Enter the amount of time to wait for
communication time-out with a temperature controller before reporting
an error.

Serial Port: Select the serial port where you will connect the
temperature controller.

78 MTS Series 793 Utility Software

HWI Hardware Settings