Metrohm IC Net 2.0 User Manual

Page 110

Advertising
background image

7 Methods

IC Net 2.0

102

Internal standard amount

Concentration of the internal standard

component for relative concentration cal-

culations.

Date/time when...

Date and time of sample collection (the

default values are equal to the date and

time when the chromatogram starts).

7.2.3

Column

Tab

Column

of passport with column information.

Number

Serial number of column (max. 256 char-

acters).

ID

Internal diameter of the column in mm.

Length

Length of the column in mm. This parame-

ter is used to calculate

Linear flow

.

Packing material

Column

Column description (max. 256 characters).

Particle size

Particle size of the column in

µ

m. This value

is used for the calculation of reduced

theoretical plate height

Reduced TP height

.

Void volume

Void volume for the column in %. Used to

calculate

Logarithmic index

,

Capacity

factor

and

Linear flow

. It is calculated by the sys-

tem in accordance with the settings defined

on

Method setup / Math

.

Precolumn

ID

Internal diameter of the precolumn in mm.

Length

Length of the precolumn in mm (set length

to 0 if no precolumn is used).

Advertising