7 Methods

IC Net 2.0

144

type

The types of components are designated

by a one-letter code:

R

Reference component (used for peak iden-

tification).

S

Standard component.

C

Calibration standard in the case it is differ-

ent from the quantification standard.

?

Some points of corresponding peak are out

of ADC or detector range – suspicious re-

sult.

!

Component concentration is outside of

minimum (

min C

) and maximum limits

(

max C

), set in the component table.

p

Special component (it has one of the spe-

cial flags checked on in the component

window).

N

The component response is outside of

calibrated region.
In the case of peaks additional letters pre-

cede the component type information, indi-

cating, how peaks are separated from other

peaks, for example:

BD_

Peak that starts on the baseline (

B

) and

ends on the drop line (

D

) that separates it

from another adjacent peak.

BBR

Special case of a rider (

R

) peak that is tan-

gentially separated from the main one. A

main peak will have a third letter

H

(horse)

in this case.
The complete component type may look

like:

BBD : !R

.

group

Number of the group for the component.

spectral ratio

Ratio of responses on different channels for

multi-channel chromatograms.

name

Name of the component.

file name

File name of chromatogram. This column is

very useful for the purpose of processing of

exported data.

ident

Title (identifier) of the chromatogram. This

column is useful for the processing of ex-

ported data.

Advertising