9.3 Sample queue editor

IC Net 2.0

167

Vial

Autosampler vial position to take sample

from.

Volume

Injected volume of sample in

µ

L.

Dilution

Sample dilution prior to injection.

Amount

Sample amount. If this value is different for

the calibration run (c) and the sample run

(s), the component concentrations of the

sample are calculated as follows:

C

s

= C

c

Amount

s

/

Amount

c

Internal standard amount

Concentration of the internal standard

component for relative concentration cal-

culations.

Level

Calibration level (see section 7.5) for the

sample: Level

0

stays for normal analysis,

levels

1

and greater for calibration runs.

Correct filling of this column enables to cal-

culate calibration coefficients automatically.

Injections

Number of injections for the same vial po-

sition (only 1 injection is permissible for

Level

≥≥ 1

).

Done

Indication whether the sample determina-

tion has been done or not:

0

determination not yet done

1

…n

determination or injection done

Sample Info 1

First sample description.

Sample Info 2

Second sample description.

9.3.2

Sample queue editor functions

The functions in the editor window for sample queues can be

triggered with the corresponding menu items of the

Edit

menu or

with the corresponding symbols in the symbol bar.

QUEUE EDITOR /

Edit / Undo

Undo the last modification of the sample queue table.

QUEUE EDITOR /

Edit / Cut row(s)

Cut the selected rows of the sample queue table and copy them

into the clipboard.

QUEUE EDITOR /

Edit / Copy row(s)

Copy the selected rows of the sample queue table into the clip-

board.