10.2 Perform batch reprocessing

IC Net 2.0

171

Use method from file for reprocessing

Selection of the desired chromatogram file

whose method should be used for reproc-

essing.

If the reprocessing includes reintegration
and/or recalibration, only chromatograms
recorded with the same method should

be loaded into the batch reprocessing
queue.

<Open example>

Open the chromatogram selected above in

the

Use method...

field.

<Open all files>

Open all chromatograms of the batch re-

processing queue.

<Edit sample table>

Open the batch reprocessing editor pro-

gram for editing the batch reprocessing ta-

ble (see section 10.3).

Reprocess sample runs

Reprocess all sample chromatograms

(

calibration level = 0

).

Reprocess calibration runs

Reprocess all calibration chromatograms

(

calibration level > 0

).

Update method file in <METHODS> directory after reprocessing

Save the method file

*.mtw

after reprocess-

ing if the method is changed.

Reintegrate

Reintegrate chromatograms according to

the current settings of the integration pa-

rameters and integration events.

<Edit integration parameters>

Open the

Integration parameters

window for

modification of integration parameters and

integration events.

Recalibrate

Reprocess all calibration chromatograms (if

Reprocess calibration runs

is switched on)

and apply new calibration to all sample

chromatograms (if

Reprocess sample runs

is

switched on) by updating the concentration

table.

Default scheme

Default setting for recalibration reprocess-

ing. The two options

Apply final calibration...

and

Forget calibration points...

are switched

on. A new calibration is performed with the

calibration runs and the resulting new cali-