Calibration and quantification, General information, Notations – Metrohm IC Net 2.0 User Manual

Page 127: External standard calibration, Component table, Peak identification, Concentration table, Calibration curve, Update calibration, Calibration data handling

Advertising
background image

7.5 Calibration and quantification

IC Net 2.0

119

7.5

Calibration and quantification

7.5.1

General information

The aim of any chromatographic analysis is to answer the ques-

tion "What components are present in the sample and what are

their concentrations?". Two procedures are used to achieve this

goal: the first step is called calibration, the second step includes

quantification.

Calibration has two aims: to get retention characteristics for all

components of interest (these data are stored in the component

table) and to establish a relation between injected amounts and

corresponding instrumental responses for all components of in-

terest (stored in the concentration table). Calibration is performed

by running one or several chromatograms of samples with known

composition and known concentration of components

(standards). For each calibrated component a calibration curve

is constructed as a result.

With «IC Net» three different procedures can be used for the con-

struction of the calibration curve. By far the most important

method for ion chromatography is the external standard cali-

bration (absolute calibration) which is described in detail in this

section. The other methods of internal standard calibration

(relative calibration) and tabulated calibration (relative gradient

factor, a modified method for external standard calibration) are of

lesser importance and are not described in detail here (for details

please refer to on-line help).

Identification is a procedure that enables to decide what peaks

on the chromatogram correspond to what components. The

identification is performed on the basis of the

Component table

created for calibration.

Quantification is a calculation procedure that determines com-

ponents concentrations, on the basis of instrumental response

(peak height or area), using the calibration curves obtained ear-

lier for each component.

Advertising