10 batch reprocessing, 1 batch reprocessing queue file handling, 1 open batch reprocessing queue – Metrohm IC Net 2.0 User Manual

Page 177: 2 create new batch reprocessing queue, 3 save batch reprocessing queue

Advertising
background image

10.1 Batch reprocessing queue file handling

IC Net 2.0

169

10 Batch reprocessing

Reprocessing is understood to be the subsequent reprocessing

of a series of chromatograms which have been loaded in a
queue (Batch reprocessing queue). For reprocessing ac-

cording to a selected method the settings for calibration, integra-

tion, passport, appearance and report can be altered at will.
A batch reprocessing queue is stored in a batch reprocessing

queue file

*.bar

in the

Data

directory.

10.1

Batch reprocessing queue file handling

10.1.1 Open batch reprocessing queue

IC NET / File / Open / Batch reprocessing

Load an existing batch reprocessing file (

*.bar

) from the

Data

di-

rectory and open the

Reprocess

window.

IC NET / File / Open / Last batch

Load the last opened batch reprocessing file (

*.bar

) from the

Data

directory and open the

Reprocess

window.

10.1.2 Create new batch reprocessing queue

IC NET / File / Open / Chromatogram

For creation of a new batch reprocessing queue, open the

Chro-

matogram open

window. Select the desired chromatograms (

*.chw

)

and click the

<To Batch>

button. Enter the name of the new batch

reprocessing file and press

<OK>

. The selected chromatograms

are then added to this batch reprocessing queue.

If the reprocessing includes reintegration and/or recalibration,
only chromatograms recorded with the same method should be

loaded into the batch reprocessing queue.

Advertising