Metrohm IC Net 2.0 User Manual

Page 51

Advertising
background image

5.4 761 Compact IC

IC Net 2.0

43

IC pump

High-pressure pump.

<On>

Start high-pressure pump.

<Off>

Stop high-pressure pump.

Peristaltic pump

Peristaltic pump.

<On>

Start peristaltic pump.

<Off>

Stop peristaltic pump.

System startup values

The system startup values are sent and

applied to the 761 Compact IC each time

the system is connected, a determination is

started, or the values are sent manually

with

<Send to unit>

.

Flow, mL/min

Startup value for flow rate of the high-

pressure pump.

Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum pressure limit

for high-pressure pump. This limit is con-

trolled even without connection to the PC.

Entry range:

0.0

...

25.0 MPa

Pressure min, MPa

Startup value for minimum pressure limit for

high-pressure pump. This limit is controlled

even without connection to the PC.

Entry range:

0.0

...

25.0

MPa

Full scale,

µµS/cm

Startup value for full scale range.

Selection:

50

,

250

,

1000

µµS/cm

Remote lines

Startup value for remote line settings 1

8.

Selection:

0

,

1

Initial

The

Initial

tab of the

761 Compact IC

window for settings of system

startup values is only available for a disconnected system.

Flow, mL/min

Startup value for flow rate of the high-

pressure pump.

Entry range:

0.20

...

2.50 mL/min