4.5 System state window

IC Net 2.0

27

Running

Running program or chromatogram data

acquisition.

Running program

(xxx min left)

Running time program with time display.

Waiting for INJECT

Waiting for "INJECT" to start program

and/or chromatogram data acquisition as

defined in the Start mode.

INJECT done

Injection valve has been switched to the

"INJECT" position.

Stopping

Determination has been stopped.

Finished

Program or chromatogram data acquisition

has been finished.

SHUTDOWN

System is shutdown.

4.5.2

General error messages

Detection of hard-

ware failed

Bad connection between PC and device or

device switched off (check connecting ca-

ble or switch on instrument).

4.5.3

Error messages for 761 Compact IC

LEAK DETECTED

The leak detector has detected a leak

(check IC system and connections).

E1

Program checksum wrong (call Metrohm

service).

E2

RAM faulty (call Metrohm service).

E200

Invalid instrument adjustment (call Metrohm

service).

E237

Storage of configuration values failed

(repeat last action; if error reappears, call

Metrohm service).

E238

Storage of instrument number failed (repeat

last action; if error reappears, call Metrohm

service).

E240

EEPROM faulty (call Metrohm service).

E258

Storage of setup values failed (repeat last

action; if error reappears, call Metrohm

service).