IC Net 2.0

137

7.6.4

Report destination

The following output devices can be selected simultaneously in

any combination as targets for report output:

Screen

Output report (without curves) to screen.

Printer

Output report to printer.

File

Output report to file (see section 7.6.7).

7.6.5

Peak table

The following parameters can be selected for peak table output

with determination results:

Quantification method

Selection of the method to calculate con-

centrations of components:

Response normalization

The method normalizes the sum of re-

sponses for all peaks of the chromatogram

to the NORM factor entered in the

Total %

for normalization

field:

R

i

% = NORM

R

i

/

д

д

R

i

.

All peaks are calculated and reported no

matter whether they have been recognized

or not.
The peak table includes the following col-

umns:

Peak number

,

Retention time

,

Area +

Area% or Height + Height%

,

Name

Normalized concentration

The method normalizes the sum of abso-

lute concentrations for all peaks of the

chromatogram to the NORM factor entered

in the

Total % for normalization

field:

C

i

% = NORM

W

i

(R

i

) /

д

д

W

i

(R

i

)

As long as no universal component is de-

fined, only peaks with non-zero concentra-

tions are listed.
The peak table includes the following col-

umns:

Peak number

,

Retention time

,

Height

,

Area

,

Response factor

,

Concentration%

,

Name

Absolute concentration

This quantification method reports raw

quantity (absolute concentration) of the

component which is calculated directly

from the calibration formula:

Q

i

= W

i

(R

i

) / V'