7.5 Calibration and quantification

IC Net 2.0

129

Base

Base for calculations (

Area

or

Height

) in

quantification and calibration procedures.

Indicates, which parameter is to be used as

a peak response.

Reference channel

Channel which is used to measure area or

height (read-only).

Formula

Formula for calibration function. There are

six possible calibration dependencies for

linear and nonlinear calibration curves.

Statistical weight

Weighting parameter used to weight cali-

bration points deviations on calculating co-

efficients of the calibration formula. Apart

from the uniform weighting

1

the values

1/x

,

1/x

2

,

1/y

and

1/y

2

are also possible.

Standard component

Name of standard component in the case

of

Internal standard calibration

or

Tabulated

calibration

method (details see on-line

help).

Concentration of standard

Concentration of standard component in

the case of

Internal standard calibration

or

Tabulated calibration

method (details see on-

line help).

7.5.8

Update calibration

IC NET / Method / Calibration / Update

Selection of this menu item opens the

Update

window.

Retention time

Replace the expected retention times of the

components with the retention times of the

corresponding peaks obtained in the cur-

rent run.

Indexes

Fill

Index

field for components with zero in-

dex.

Coefficients

Recalculate all calibration coefficients for all

components (performs recalibration).